On-Line Control of a Styrene Tower with a ClearView® db Photometer

Purpose:

Guided Wave's ClearView db photometer can replace on-line GCs for rapid control of a styrene purification column by measuring styrene at two points.

Background:

The first sample point is from the over- head with 0 to 80% styrene in ethyl- benzene, but also containing up to 6% toluene and 2% benzene. The second is at a middle tray with 60 to 100% styrene in ethylbenzene with toluene and benzene stripped out.

Approach:

Samples at 40 oC in a 10mm cuvette were run on our diode array spectrophotometer to identify the relevant wavelengths. (See Figure 1). Two NIR wavelengths (down arrows) were required for the more complicated mixture in the overhead line, and only one for the middle tray product stream. The samples were then re-run in the photometer, absorbance value was recorded, and the regression analyses were performed.

Results:

The MLR calibration for styrene in 10 overhead samples resulted in an R2of 0.999 with a standard error of 0.84% styrene. The predicted vs. actual plot for this model is shown in Figure 2. The styrene calibration for the simpler stream
resulted in the same R2 and a standard error of 0.37%. The photometer was in- stalled on line with two side stream flow probes using fiber optic cables and the factory calibrations. The results of over three months of data for the overhead stream is shown in Figure 3 compared with laboratory values for grab samples using a gas chromatograph. The average difference is 0.6 with a standard deviation of 1.2. The simpler stream was sampled infrequently because the analyzer’s styrene values matched the GC values within experimental error.

Conclusion:

A ClearView db can provide real-time, online measurements of both streams with no system down-time, permitting greater column efficiency. For the example above ROI was achieved in less than 6 months.

Clearview db Enclourse Options Figure 3
On-Line Control of a Styrene Tower with a ClearView® db Figure 1
Residual Analysis Figure 2
Styrene in Colum Overhead Figure 3